VWR® Incubating Orbital Shaker, Model 5000I

Supplier: VWR International

Shaker offers a diverse combination of advanced operating and safety features for reliable, high-performance shaking and incubating applications in cell cultures, solubility studies, bacterial suspensions, and extraction procedures.

  • Exclusive opti-flow forced ventilation system
  • Exclusive Accu-Drive™ shaking system
  • Now with Best-in-Class, Exclusive 5-Year Warranty
  • Available with a NIST Traceable, ISO/IEC 17025 Accredited Calibration Certificate from Troemner
Variable speed microprocessor control provides consistent and uniform mixing action with the ability to restart and display last set-point if power is interrupted. Precise temperature control is maintained by a PID controller, and the unit is driven by a triple eccentric drive. Permanently lubricated bearings and a maintenance-free brushless DC motor ensure reliable service and continuous duty operation. Shaker is equipped with the Accu-Drive™ shaking system, which delivers exceptional speed control, accuracy, and durability. The system continuously monitors shaking speed and maintains setpoint even under changing loads. The displayed speed is accurate to 1% of setting. The Opti-Flow ventilation system utilizes twin fans and air deflectors to deliver optimal temperature uniformity. Built-in temperature calibration mode and speed calibration mode allow users to calibrate temperature to an external device, or to automatically recalibrate the speed display.

A variety of safety features are standard on the Model 5000I shaker. A built-in load sensor detects unbalanced conditions and reduces rpm or stops the unit to protect samples. Audible and visual signals will activate if the system detects a tray obstruction or overload. To avoid sample loss, a ramping feature slowly increases speed to setpoint. In timed mode, an alarm will sound when the timer reaches the zero point. A polycarbonate lid permits viewing of samples without disturbing internal temperature. When the lid is opened, an automatic shutoff circuit will stop the platform operation. Each unit's spill-resistant design protects internal components from accidental spills and leaks. Tray and rubber mat are removable to allow for easy chamber cleaning. A caution symbol will illuminate when air temperature in the chamber reaches 40°C (104°F). This symbol remains lit until the chamber temperature cools.

The touchpad control features easy-to-read, independent LEDs for temperature, time, and speed. An RS-232 interface provides two-way connection for data logging and unit control.

Ordering Information: Shaker includes a non-skid rubber mat that fits the 45.7x45.7 cm (18x18") tray. Tray is suitable for holding Petri dishes and cell culture flasks. Units also includes a 233 cm (92") detachable 3-wire cord and plug. Optional stainless steel platform can be used for mounting Erlenmeyer flask clamps and test tube racks. 230V, 50/60Hz models are also available; contact your VWR sales representative for more information.

RoHS certified. TUV listed. Calibrated Shakers include a Troemner ISO/IEC Accredited Calibration Certificate traceable to NIST. The ISO/IEC 17025 Accredited Calibration Certificate includes multiple data points within the speed range and time function with the associated uncertainties. Troemner is accredited by NVLAP under Laboratory Code 105013-0 to perform ISO/IEC 17025 calibrations.



Log in to see your contract pricing and availability.
  Remember me on this device
 

ORDER

Expand All / Collapse All

Shakers
Description Electrical Orbit Speed VWR Catalog Number Unit Price Quantity
Model 5000I Shaker 120V 25.4 mm (1") 15–500 rpm 97009-890 Each Retrieving

Shakers with NIST Traceable Certificate
Description Electrical Orbit Speed VWR Catalog Number Unit Price Quantity
Model 5000I Shaker 120V 25.4 mm (1") 15–500 rpm 10027-258 Each Retrieving
Generate Barcodes Barcode Label Format: 
Generate PDF Catalog Page

SPECIFICATIONS

Exterior Dimensions 81.5L x 67.5W x 60H cm (32 x 261/2 x 235/8")
Interior dimensions WxDxH (mm) 52.3L x 63W x 43.2H cm (2019/32 x 2413/16 x 17")
Max. load 22.7 kg (50 lbs.)
Shipping Weight 97.5 kg (215 lbs.)
Speed accuracy ±1% of set speed, <100 rpm ± 1 rpm ±1% of set speed, \<100 rpm ± 1 rpm
Temperature Range 5° above ambient to 65°C (149°F)
Temperature Uniformity ±0.5° at 37°C (98°F)
Timer Range 1 sec. to 160 hrs.
Tray Dimensions 45.7 x 45.7 cm (18 x 18")
Enter Details

Select the article for which you are looking for Certificate

Indicate the Lot Number. This field is mandatory.

Enter Certificate Details

Select Product:

Enter the Lot number:

Choose from recent batch/lot numbers:

Or
Review & Compare Alternatives
We found alternative products that can save you up to  per item-unit.  To compare product details, select up to 3 alternatives below and click Compare Selected. To add items to your basket, enter a quantity and click Add to Basket.

Original Product:

Description Catalog Number Availability Unit Your Price Price Per Qty